High Performance Liquid Chromatograph (HPLC) System
Brand
Waters
Model
Arc HPLC
Place of origin
USA
Location
NE3-302
Main specifications
1. Gradient Mode: Low-pressure mixing, quaternary gradient, allowing arbitrary mixing of 1-4 solvent channels
2. Flow Rate Range: 0.001-5.000 mL/min with 0.001 mL/min increments
3. Maximum Operating Pressure: 9,500 psi (655 bar)
4. Flow Precision: <0.075% RSD error
5. Flow Accuracy: ±1.0% at flow rates of 0.5, 3.0 and 5.0 mL/min
6. Gradient Profile Options: Programmable gradient curves including:
1 linear type
2 step types
4 concave curves
4 convex curves
(Total: 11 curve types across 4 categories)
7. Sample Compartment Temperature Range: 4°C-40°C with 0.1°C increments
8. Column Capacity: Single column configuration; Maximum inner diameter: 7.8 mm; Maximum length: 300 mm
Main functions
- The system effectively detects trace pollutants in complex environmental samples, including organic components in atmospheric particulates and aerosols such as polycyclic aromatic hydrocarbons (PAHs), organic acids, aldehydes & ketones, and secondary organic aerosol (SOA) markers;
- Its applications span atmospheric chemistry research, environmental monitoring, and climate change studies - for instance, analyzing organic components in PM2.5/PM10 to investigate pollutant transport and transformation pathways, or combining with isotope-ratio analysis to distinguish contributions from biogenic and anthropogenic emissions;
- The system is equipped with high-performance optical detectors including a photodiode array detector (PDA) and UV/Vis detector